Tactile sensors are widely employed to enable the sense of touch for applications such as robotics and prosthetics. In addition to the selection of an appropriate sensing material, the performance of the tactile sensing system is conditioned by its interface electronic system. On the other hand, due to the need to embed the tactile sensing system into a prosthetic device, strict requirements such as small size and low power consumption are imposed on the system design. This paper presents the experimental assessment and characterization of an interface electronic system for piezoelectric tactile sensors for prosthetic applications. The interface electronic is proposed as part of a wearable system intended to be integrated into an upper limb prosthetic device. The system is based on a low power arm-microcontroller and a DDC232 device. Electrical and electromechanical setups have been implemented to assess the response of the interface electronic with PVDF-based piezoelectric sensors. The results of electrical and electromechanical tests validate the correct functionality of the proposed system.
Experimental assessment of the interface electronic system for PVDF-based piezoelectric tactile sensors
Yahya Abbass;Ali Ibrahim;Maurizio Valle
2019-01-01
Abstract
Tactile sensors are widely employed to enable the sense of touch for applications such as robotics and prosthetics. In addition to the selection of an appropriate sensing material, the performance of the tactile sensing system is conditioned by its interface electronic system. On the other hand, due to the need to embed the tactile sensing system into a prosthetic device, strict requirements such as small size and low power consumption are imposed on the system design. This paper presents the experimental assessment and characterization of an interface electronic system for piezoelectric tactile sensors for prosthetic applications. The interface electronic is proposed as part of a wearable system intended to be integrated into an upper limb prosthetic device. The system is based on a low power arm-microcontroller and a DDC232 device. Electrical and electromechanical setups have been implemented to assess the response of the interface electronic with PVDF-based piezoelectric sensors. The results of electrical and electromechanical tests validate the correct functionality of the proposed system.File | Dimensione | Formato | |
---|---|---|---|
sensors-19-04437-v2.pdf
accesso aperto
Descrizione: file in post print
Tipologia:
Documento in versione editoriale
Dimensione
4.5 MB
Formato
Adobe PDF
|
4.5 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.