Purpose: To evaluate the accuracy and precision of a digital scanner used to scan four implants positioned according to an immediate loading implant protocol and to assess the accuracy of an aluminum framework fabricated from a digital impression. Materials and Methods: Five master casts reproducing different edentulous maxillae with four tilted implants were used. Four scan bodies were screwed onto the low-profile abutments, and a digital intraoral scanner was used to perform five digital impressions of each master cast. To assess trueness, a metal framework of the best digital impression was produced with computeraided design/computer-assisted manufacture (CAD/CAM) technology and passive fit was assessed with the Sheffield test. Gaps between the frameworks and the implant analogs were measured with a stereomicroscope. To assess precision, three-dimensional (3D) point cloud processing software was used to measure the deviations between the five digital impressions of each cast by producing a color map. The deviation values were grouped in three classes, and differences were assessed between class 2 (representing lower discrepancies) and the assembled classes 1 and 3 (representing the higher negative and positive discrepancies, respectively). Results: The frameworks showed a mean gap of < 30 μm (range: 2 to 47 μm). A statistically significant difference was found between the two groups by the 3D point cloud software, with higher frequencies of points in class 2 than in grouped classes 1 and 3 (P < .001). Conclusion: Within the limits of this in vitro study, it appears that a digital impression may represent a reliable method for fabricating full-arch implant frameworks with good passive fit when tilted implants are present.

Precision and Accuracy of a Digital Impression Scanner in Full-Arch Implant Rehabilitation

Pesce, Paolo;Setti, Paolo;Menini, Maria
2018-01-01

Abstract

Purpose: To evaluate the accuracy and precision of a digital scanner used to scan four implants positioned according to an immediate loading implant protocol and to assess the accuracy of an aluminum framework fabricated from a digital impression. Materials and Methods: Five master casts reproducing different edentulous maxillae with four tilted implants were used. Four scan bodies were screwed onto the low-profile abutments, and a digital intraoral scanner was used to perform five digital impressions of each master cast. To assess trueness, a metal framework of the best digital impression was produced with computeraided design/computer-assisted manufacture (CAD/CAM) technology and passive fit was assessed with the Sheffield test. Gaps between the frameworks and the implant analogs were measured with a stereomicroscope. To assess precision, three-dimensional (3D) point cloud processing software was used to measure the deviations between the five digital impressions of each cast by producing a color map. The deviation values were grouped in three classes, and differences were assessed between class 2 (representing lower discrepancies) and the assembled classes 1 and 3 (representing the higher negative and positive discrepancies, respectively). Results: The frameworks showed a mean gap of < 30 μm (range: 2 to 47 μm). A statistically significant difference was found between the two groups by the 3D point cloud software, with higher frequencies of points in class 2 than in grouped classes 1 and 3 (P < .001). Conclusion: Within the limits of this in vitro study, it appears that a digital impression may represent a reliable method for fabricating full-arch implant frameworks with good passive fit when tilted implants are present.
File in questo prodotto:
File Dimensione Formato  
Pesce 2018 impression.pdf

accesso chiuso

Tipologia: Documento in versione editoriale
Dimensione 201.67 kB
Formato Adobe PDF
201.67 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11567/895308
Citazioni
  • ???jsp.display-item.citation.pmc??? 23
  • Scopus 79
  • ???jsp.display-item.citation.isi??? 73
social impact