The purpose of this report is to describe deployment of the Relay NBS Thoracic Stent Graft with the Plus Delivery System (Bolton Medical, Sunrise, FL) in a flexible resin arch model with a 15-mm radius curve as well as our preliminary clinical results. The Relay NBS graft with the Plus Delivery System was evaluated by way of bench testing, which was performed with stent grafts with diameters ranging from 24 to 46 mm and lengths ranging from 100 to 250 mm in flexible resin arch models with a 15-mm arch radius of curvature. The deployment sequence was analyzed. The Relay NBS graft with the Plus Delivery System was deployed in two patients, respectively, having a 6.5-cm penetrating aortic ulcer of the proximal third of the descending thoracic aorta and a DeBakey type-I aortic dissection with chronic false lumen dilatation after surgery due to an entry site at the distal thoracic aorta. Bench tests showed proper conformation and apposition of the Relay NBS graft with the Plus Delivery System in the flexible resin model. This stent graft was deployed successfully into the two patients with a correct orientation of the first stent and without early or late complications. The Relay NBS graft with the Plus Delivery System ensures an optimal conformation and apposition of the first stent in the aortic arch with a small radius of curvature.

Relay NBS Graft with the Plus Delivery System to Improve Deployment in Aortic Arch with Small Radius Curve

SCARANO, FLAVIO;PASSERONE, GIANCARLO
2011-01-01

Abstract

The purpose of this report is to describe deployment of the Relay NBS Thoracic Stent Graft with the Plus Delivery System (Bolton Medical, Sunrise, FL) in a flexible resin arch model with a 15-mm radius curve as well as our preliminary clinical results. The Relay NBS graft with the Plus Delivery System was evaluated by way of bench testing, which was performed with stent grafts with diameters ranging from 24 to 46 mm and lengths ranging from 100 to 250 mm in flexible resin arch models with a 15-mm arch radius of curvature. The deployment sequence was analyzed. The Relay NBS graft with the Plus Delivery System was deployed in two patients, respectively, having a 6.5-cm penetrating aortic ulcer of the proximal third of the descending thoracic aorta and a DeBakey type-I aortic dissection with chronic false lumen dilatation after surgery due to an entry site at the distal thoracic aorta. Bench tests showed proper conformation and apposition of the Relay NBS graft with the Plus Delivery System in the flexible resin model. This stent graft was deployed successfully into the two patients with a correct orientation of the first stent and without early or late complications. The Relay NBS graft with the Plus Delivery System ensures an optimal conformation and apposition of the first stent in the aortic arch with a small radius of curvature.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11567/566119
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 10
social impact