The selection of the photomultiplier plays a crucial role in the R&D activity related to a large-scale underwater neutrino telescope. This paper illustrates the main procedures and facilities used to characterize the performances of 72 large area photomultipliers, Hamamatsu model R7081 sel. The voltage to achieve a gain of 5 x 107, dark count rate and single photoelectron time and charge properties of the overall response were measured with a properly attenuated 410 nm pulsed laser. A dedicated study of the spurious pulses was also performed. The results prove that the photomultipliers comply with the general requirements imposed by the project.
Procedures and results of the measurements on large area photomultipliers for the NEMO project
FRATINI, KATIA;RICCO, GIOVANNI;TAIUTI, MAURO GINO;
2010-01-01
Abstract
The selection of the photomultiplier plays a crucial role in the R&D activity related to a large-scale underwater neutrino telescope. This paper illustrates the main procedures and facilities used to characterize the performances of 72 large area photomultipliers, Hamamatsu model R7081 sel. The voltage to achieve a gain of 5 x 107, dark count rate and single photoelectron time and charge properties of the overall response were measured with a properly attenuated 410 nm pulsed laser. A dedicated study of the spurious pulses was also performed. The results prove that the photomultipliers comply with the general requirements imposed by the project.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.