The Authors reported investigations and results of a twenty-year experience with femoral head/socket coupling in allumina ceramic related to 248 total hip prostheses which were implanted between 1983 and 1999, and periodically controlled. At a follow-up in December 2003, 161 implants were lost and only 28 implants survived and were checked with a mean follow-up of 15.5 years (minimun 10 years,maximum 20 years). The high rate of failure, especially cup failure, was correlated to poor features of the Biodirecta prosthetic design: in particular, the alumina cup was too stiff in respect to the bone.
20 ANNI DI ESPERIENZA CON L'ACCOPPIAMENTO ALLUMINA-ALLUMINA NELLE PROTESI D'ANCA
FRANCHIN, FRANCESCO MARIA;SANGUINETI, FRANCESCA;
2004-01-01
Abstract
The Authors reported investigations and results of a twenty-year experience with femoral head/socket coupling in allumina ceramic related to 248 total hip prostheses which were implanted between 1983 and 1999, and periodically controlled. At a follow-up in December 2003, 161 implants were lost and only 28 implants survived and were checked with a mean follow-up of 15.5 years (minimun 10 years,maximum 20 years). The high rate of failure, especially cup failure, was correlated to poor features of the Biodirecta prosthetic design: in particular, the alumina cup was too stiff in respect to the bone.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.