Objectives To compare torque recordings of immediately loaded orthodontic miniscrews between insertion time and different post-placement timepoints (2 weeks, 4 weeks and removal time, respectively).Setting and sample population Parallel trial with an allocation ratio of 1:1. Eligibility criteria were needs of fixed orthodontic treatment, no systemic disease and absence of using drugs altering bone metabolism.Material and methods Patients received miniscrews, 2.0 mm diameter and 10 mm length. All miniscrews underwent inter-radicular placement, and they were placed in the maxilla or in the mandible, palatally or buccally. No pre-drilling was performed. Miniscrews were loaded immediately after the insertion and were used for distalization, intrusion, extrusion, mesialization or indirect anchorage. Patients were randomly divided into three groups. For each patient, Maximum Insertion Torque (MIT) was evaluated at baseline. MIT was measured again after 2 weeks and after 4 weeks by tightening the screw a quarter of turn in Groups 1 and 2, respectively. At the end of the treatment, maximal removal torque was evaluated in Group 3. Torque variation with respect to insertion time was considered as the primary outcome. Baseline and longitudinal differences were tested using the linear mixed-effects (LME) model.Results Forty seven patients and 74 miniscrews were followed up. An association existed between maximum insertion torque and the observation time. A torque decrease of 26.9% and 30% after 2 weeks was observed for mandibular and maxillary miniscrews, respectively. After 1 month, torque values were similar to the baseline records. The overall success rate was 79.7%. No serious harm was observed.Conclusions Maximum insertion torque undergoes a loss during the first 2 weeks, and its values may depend on the insertion site and the anchorage purpose. Removal torque value is almost the same as the initial torque after 1 month.

Maximum insertion torque loss after miniscrew placement in orthodontic patients: A randomized controlled trial

Migliorati, Marco;Drago, Sara;Silvestrini-Biavati, Armando
2021-01-01

Abstract

Objectives To compare torque recordings of immediately loaded orthodontic miniscrews between insertion time and different post-placement timepoints (2 weeks, 4 weeks and removal time, respectively).Setting and sample population Parallel trial with an allocation ratio of 1:1. Eligibility criteria were needs of fixed orthodontic treatment, no systemic disease and absence of using drugs altering bone metabolism.Material and methods Patients received miniscrews, 2.0 mm diameter and 10 mm length. All miniscrews underwent inter-radicular placement, and they were placed in the maxilla or in the mandible, palatally or buccally. No pre-drilling was performed. Miniscrews were loaded immediately after the insertion and were used for distalization, intrusion, extrusion, mesialization or indirect anchorage. Patients were randomly divided into three groups. For each patient, Maximum Insertion Torque (MIT) was evaluated at baseline. MIT was measured again after 2 weeks and after 4 weeks by tightening the screw a quarter of turn in Groups 1 and 2, respectively. At the end of the treatment, maximal removal torque was evaluated in Group 3. Torque variation with respect to insertion time was considered as the primary outcome. Baseline and longitudinal differences were tested using the linear mixed-effects (LME) model.Results Forty seven patients and 74 miniscrews were followed up. An association existed between maximum insertion torque and the observation time. A torque decrease of 26.9% and 30% after 2 weeks was observed for mandibular and maxillary miniscrews, respectively. After 1 month, torque values were similar to the baseline records. The overall success rate was 79.7%. No serious harm was observed.Conclusions Maximum insertion torque undergoes a loss during the first 2 weeks, and its values may depend on the insertion site and the anchorage purpose. Removal torque value is almost the same as the initial torque after 1 month.
File in questo prodotto:
File Dimensione Formato  
ocr.12449 (1).pdf

accesso aperto

Tipologia: Documento in Post-print
Dimensione 943.65 kB
Formato Adobe PDF
943.65 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11567/1121518
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact