We implemented a laboratory apparatus based on a NaI(TI) γ-ray spectrometer for the assessment of naturally occurring radioactive materials. This apparatus is devoted to the quantitative determination of U, Th and K in rock and soil samples. We analyzed the calibration procedures with particular emphasis to factors which affect accuracy and detection limits. A first series of measurements were compared to those obtained by means of high energy γ-ray spectrometry. A discussion was made on factors restricting the approach based on scintillation spectrometry, such as secular equilibrium and radon encapsulation.

Laboratory application of NaI(Tl) γ-ray spectrometry to studies of natural radioactivity in geophysics

Chiozzi P.;Verdoya M.
2000-01-01

Abstract

We implemented a laboratory apparatus based on a NaI(TI) γ-ray spectrometer for the assessment of naturally occurring radioactive materials. This apparatus is devoted to the quantitative determination of U, Th and K in rock and soil samples. We analyzed the calibration procedures with particular emphasis to factors which affect accuracy and detection limits. A first series of measurements were compared to those obtained by means of high energy γ-ray spectrometry. A discussion was made on factors restricting the approach based on scintillation spectrometry, such as secular equilibrium and radon encapsulation.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11567/1108081
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 82
  • ???jsp.display-item.citation.isi??? 79
social impact