Objectives: In this study we investigated the rate of susceptibility testing discrepancies between semi-automated and reference systems with carbapenem-resistant Enterobacterales (CRE) and the impact of alleged errors by semi-automated systems on guiding targeted therapy for CRE bloodstream infection (BSI). Methods: This was a multicentre, retrospective study enrolling patients with monomicrobial BSI caused by CRE from January 2013 to December 2016. Nonduplicate isolates from index blood cultures tested locally with semi-automated systems were centralized at a referral laboratory and retested with a reference broth microdilution or agar dilution method. Results: We enrolled 366 patients with CRE-BSI; 220 (60%) were male, and the median age was 67 years (interquartile range, 54–76 years). When compared with the results of the reference methods, those of the semi-automated systems exhibited variable rates of very major errors (VMEs; i.e. false susceptibilities) and major errors (MEs; i.e. false resistances). The highest rates of VMEs were observed with fosfomycin (14%) and colistin (13.9%), and the highest rates of MEs were observed with gentamicin (21%), fosfomycin (7.7%), and tigecycline (34%). Overall, VMEs and MEs led clinicians to prescribe or confirm ineffective therapy in 25 of 341 patients (7%). Receipt of ineffective therapy supported by a misleading susceptibility test was associated with higher 30-day mortality rates by Kaplan–Meier survival curves rates compared with receipt of active therapy (56% vs. 26%; p = 0.002), and the difference was confirmed after adjustment for confounders in a Cox regression model (adjusted hazard ratio: 2.91; 95% CI, 1.62–5.22; p < 0.001). Discussion: MEs and VMEs were relatively common with semi-automated susceptibility testing systems. VMEs were associated with inappropriate use of antibiotics and poorer outcomes.

Clinical consequences of very major errors with semi-automated testing systems for antimicrobial susceptibility of carbapenem-resistant Enterobacterales

Giacobbe D. R.;Pascale R.;Dettori S.;Bassetti M.;De Rosa F. G.;Marchese A.;Cavallo R.;Viale P.;
2022-01-01

Abstract

Objectives: In this study we investigated the rate of susceptibility testing discrepancies between semi-automated and reference systems with carbapenem-resistant Enterobacterales (CRE) and the impact of alleged errors by semi-automated systems on guiding targeted therapy for CRE bloodstream infection (BSI). Methods: This was a multicentre, retrospective study enrolling patients with monomicrobial BSI caused by CRE from January 2013 to December 2016. Nonduplicate isolates from index blood cultures tested locally with semi-automated systems were centralized at a referral laboratory and retested with a reference broth microdilution or agar dilution method. Results: We enrolled 366 patients with CRE-BSI; 220 (60%) were male, and the median age was 67 years (interquartile range, 54–76 years). When compared with the results of the reference methods, those of the semi-automated systems exhibited variable rates of very major errors (VMEs; i.e. false susceptibilities) and major errors (MEs; i.e. false resistances). The highest rates of VMEs were observed with fosfomycin (14%) and colistin (13.9%), and the highest rates of MEs were observed with gentamicin (21%), fosfomycin (7.7%), and tigecycline (34%). Overall, VMEs and MEs led clinicians to prescribe or confirm ineffective therapy in 25 of 341 patients (7%). Receipt of ineffective therapy supported by a misleading susceptibility test was associated with higher 30-day mortality rates by Kaplan–Meier survival curves rates compared with receipt of active therapy (56% vs. 26%; p = 0.002), and the difference was confirmed after adjustment for confounders in a Cox regression model (adjusted hazard ratio: 2.91; 95% CI, 1.62–5.22; p < 0.001). Discussion: MEs and VMEs were relatively common with semi-automated susceptibility testing systems. VMEs were associated with inappropriate use of antibiotics and poorer outcomes.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11567/1098386
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 10
social impact