Purpose: To evaluate the safety, effectiveness, and the best parameters setup of hypersonic vitrectomy. Methods: A prospective, multicentric, interventional study on 50 eyes that had undergone hypersonic vitrectomy because of various vitreoretinal diseases. We primarily assessed the effectiveness of vitreous removal, intraoperative complications, and surgical setup. Secondarily, we evaluated single-surgery anatomical success and postoperative best-corrected visual acuity. Results: Intraoperative complications occurred in 5 eyes (10%), whereas technical problems were detected in 23 eyes (46%), resulting in conversion to guillotine vitrectomy in 15 cases. The most common finding related to the technical problems was an inadequate vitreous liquefaction with the formation of vitreous strands and consequent inadequate vitreous outflow, sometimes complicated by vitreous incarceration in the vitrectomy probe. The best settings were considered a stroke of 60 mm and vacuum of 40 mmHg for both core and peripheral vitrectomy. At 3-month follow-up, primary anatomical success was achieved in 49 eyes (98%) and the mean best-corrected visual acuity overall improved. Conclusion: The availability of hypersonic vitrectomy in the current surgical practice opens a new era in vitreoretinal surgery. Despite the potential advantages in fluidics, the performance of hypersonic vitrectomy system needs to be further optimized, mainly for the occurrence of inadequate vitreous liquefaction and vitreous strands formation.
Twenty-three-gauge hypersonic vitrectomy
REPETTO R.;
2021-01-01
Abstract
Purpose: To evaluate the safety, effectiveness, and the best parameters setup of hypersonic vitrectomy. Methods: A prospective, multicentric, interventional study on 50 eyes that had undergone hypersonic vitrectomy because of various vitreoretinal diseases. We primarily assessed the effectiveness of vitreous removal, intraoperative complications, and surgical setup. Secondarily, we evaluated single-surgery anatomical success and postoperative best-corrected visual acuity. Results: Intraoperative complications occurred in 5 eyes (10%), whereas technical problems were detected in 23 eyes (46%), resulting in conversion to guillotine vitrectomy in 15 cases. The most common finding related to the technical problems was an inadequate vitreous liquefaction with the formation of vitreous strands and consequent inadequate vitreous outflow, sometimes complicated by vitreous incarceration in the vitrectomy probe. The best settings were considered a stroke of 60 mm and vacuum of 40 mmHg for both core and peripheral vitrectomy. At 3-month follow-up, primary anatomical success was achieved in 49 eyes (98%) and the mean best-corrected visual acuity overall improved. Conclusion: The availability of hypersonic vitrectomy in the current surgical practice opens a new era in vitreoretinal surgery. Despite the potential advantages in fluidics, the performance of hypersonic vitrectomy system needs to be further optimized, mainly for the occurrence of inadequate vitreous liquefaction and vitreous strands formation.File | Dimensione | Formato | |
---|---|---|---|
romano-et-al-retina-2021.pdf
accesso chiuso
Descrizione: Articolo su rivista
Tipologia:
Documento in versione editoriale
Dimensione
274.9 kB
Formato
Adobe PDF
|
274.9 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.