Analysis of the microcirculation is currently performed offline, is time consuming and operator dependent. The aim of this study was to assess the ability and efficiency of the automatic analysis software CytoCamTools 1.7.12 (CC) to measure microvascular parameters in comparison with Automated Vascular Analysis (AVA) software 3.2. 22 patients admitted to the cardiothoracic intensive care unit following cardiac surgery were prospectively enrolled. Sublingual microcirculatory videos were analysed using AVA and CC software. The total vessel density (TVD) for small vessels, perfused vessel density (PVD) and proportion of perfused vessels (PPV) were calculated. Blood flow was assessed using the microvascular flow index (MFI) for AVA software and the averaged perfused speed indicator (APSI) for the CC software. The duration of the analysis was also recorded. Eighty-four videos from 22 patients were analysed. The bias between TVD-CC and TVD-AVA was 2.20 mm/mm2 (95 % CI 1.37–3.03) with limits of agreement (LOA) of −4.39 (95 % CI −5.66 to −3.16) and 8.79 (95 % CI 7.50–10.01) mm/mm2. The percentage error (PE) for TVD was ±32.2 %. TVD was positively correlated between CC and AVA (r = 0.74, p < 0.001). The bias between PVD-CC and PVD-AVA was 6.54 mm/mm2 (95 % CI 5.60–7.48) with LOA of −4.25 (95 % CI −8.48 to −0.02) and 17.34 (95 % CI 13.11–21.57) mm/mm2. The PE for PVD was ±61.2 %. PVD was positively correlated between CC and AVA (r = 0.66, p < 0.001). The median PPV-AVA was significantly higher than the median PPV-CC [97.39 % (95.25, 100 %) vs. 81.65 % (61.97, 88.99), p < 0.0001]. MFI categories cannot estimate or predict APSI values (p = 0.45). The time required for the analysis was shorter with CC than with AVA system [2′42″ (2′12″, 3′31″) vs. 16′12″ (13′38″, 17′57″), p < 0.001]. TVD is comparable between the two softwares, although faster with CC software. The values for PVD and PPV are not interchangeable given the different approach to assess microcirculatory flow.

Ability and efficiency of an automatic analysis software to measure microvascular parameters

Bazurro S.;
2017-01-01

Abstract

Analysis of the microcirculation is currently performed offline, is time consuming and operator dependent. The aim of this study was to assess the ability and efficiency of the automatic analysis software CytoCamTools 1.7.12 (CC) to measure microvascular parameters in comparison with Automated Vascular Analysis (AVA) software 3.2. 22 patients admitted to the cardiothoracic intensive care unit following cardiac surgery were prospectively enrolled. Sublingual microcirculatory videos were analysed using AVA and CC software. The total vessel density (TVD) for small vessels, perfused vessel density (PVD) and proportion of perfused vessels (PPV) were calculated. Blood flow was assessed using the microvascular flow index (MFI) for AVA software and the averaged perfused speed indicator (APSI) for the CC software. The duration of the analysis was also recorded. Eighty-four videos from 22 patients were analysed. The bias between TVD-CC and TVD-AVA was 2.20 mm/mm2 (95 % CI 1.37–3.03) with limits of agreement (LOA) of −4.39 (95 % CI −5.66 to −3.16) and 8.79 (95 % CI 7.50–10.01) mm/mm2. The percentage error (PE) for TVD was ±32.2 %. TVD was positively correlated between CC and AVA (r = 0.74, p < 0.001). The bias between PVD-CC and PVD-AVA was 6.54 mm/mm2 (95 % CI 5.60–7.48) with LOA of −4.25 (95 % CI −8.48 to −0.02) and 17.34 (95 % CI 13.11–21.57) mm/mm2. The PE for PVD was ±61.2 %. PVD was positively correlated between CC and AVA (r = 0.66, p < 0.001). The median PPV-AVA was significantly higher than the median PPV-CC [97.39 % (95.25, 100 %) vs. 81.65 % (61.97, 88.99), p < 0.0001]. MFI categories cannot estimate or predict APSI values (p = 0.45). The time required for the analysis was shorter with CC than with AVA system [2′42″ (2′12″, 3′31″) vs. 16′12″ (13′38″, 17′57″), p < 0.001]. TVD is comparable between the two softwares, although faster with CC software. The values for PVD and PPV are not interchangeable given the different approach to assess microcirculatory flow.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11567/1007027
Citazioni
  • ???jsp.display-item.citation.pmc??? 13
  • Scopus 29
  • ???jsp.display-item.citation.isi??? 23
social impact