Purpose: To assess Automatic GonioPhotography (AGP) - true color images of the anterior chamber angle (ACA) - for identifying angle anatomical structures and pathological or post-surgical findings. Methods: GS-1 (Gonioscope GS-1; NIDEK Technologies Srl) was used to acquire 360-degree images of the ACA on both eyes of 50 patients. Apparent iris insertion for each quadrant and others ACA’s findings were recorded. All patients underwent standard gonioscopy and the angle was graded using the Spaeth Classification. Inter-device agreement was determined by using Cohen's linearly weighted κ (KW) coefficient of concordance. Results: Of the 400 quadrants imaged, 339 (84.8%) were of sufficient quality for analysis. The mean (SD) patients’ age was 56 (14.2) years and 29 (58%) were women. AGP with GS-1 and standard gonioscopy showed good inter-device agreement on apparent iris insertion evaluation (KW, 0.65; 95% CI, 0.51 to 0.78). No differences between AGP and gonioscopy were observed in detecting findings such as XEN gel stent, Ex-PRESS device, long tube, trabeculectomy, anglesupported IOL. No adverse events related to gonioscopy or GS-1 examination were observed. Conclusions: AGP using the GS-1 is a reliable method for assessing apparent iris insertion and proved useful in detecting pathological and post-surgical ACA findings.

Obiettivi: valutazione della goniofotografia automatica (AGP) - immagini a colori dell’angolo camerulare (ACA) - nell’identificazione delle strutture anatomiche angolari e dei reperti patologici o post-chirurgici. Metodi: lo strumento GS-1 (Gonioscope GS-1; NIDEK Technologies Srl) è stato utilizzato per acquisire immagini a 360 gradi dell’ACA in entrambi gli occhi di 50 pazienti. L’inserzione apparente dell’iride per ogni quadrante ed eventuali altri reperti dell’ACA sono stati registrati. Tutti i pazienti sono stati sottoposti anche alla gonioscopia standard e l’angolo è stato valutato utilizzando la classificazione di Spaeth. La concordanza tra i due strumenti è stata determinata utilizzando il coefficiente di Cohen's pesato linearmente κ (KW). Risultati: dei 400 quadranti fotografati, 339 (84.8%) erano di qualità sufficiente per le analisi. L’età media (DS) era 56 (14.2) anni e 29 (58%) erano donne. La concordanza tra AGP con GS-1 e gonioscopia standard è risultata essere buona (KW, 0.65; 95% CI, 0.51 to 0.78). Non ci sono state differenze tra AGP e gonioscopia circa l’identificazione di reperti anormali quali: XEN gel stent, dispositivo Ex-PRESS, tubicino di impianto drenante, trabeculectomia, IOL a supporto angolare. Nessun evento avverso relativo alla gonioscopia o al GS-1 è stato osservato.

Clinical evaluation of a novel device for anterior chamber angle photography

Cutolo CA;Scotto R;Iester M;Traverso CE.
2018-01-01

Abstract

Purpose: To assess Automatic GonioPhotography (AGP) - true color images of the anterior chamber angle (ACA) - for identifying angle anatomical structures and pathological or post-surgical findings. Methods: GS-1 (Gonioscope GS-1; NIDEK Technologies Srl) was used to acquire 360-degree images of the ACA on both eyes of 50 patients. Apparent iris insertion for each quadrant and others ACA’s findings were recorded. All patients underwent standard gonioscopy and the angle was graded using the Spaeth Classification. Inter-device agreement was determined by using Cohen's linearly weighted κ (KW) coefficient of concordance. Results: Of the 400 quadrants imaged, 339 (84.8%) were of sufficient quality for analysis. The mean (SD) patients’ age was 56 (14.2) years and 29 (58%) were women. AGP with GS-1 and standard gonioscopy showed good inter-device agreement on apparent iris insertion evaluation (KW, 0.65; 95% CI, 0.51 to 0.78). No differences between AGP and gonioscopy were observed in detecting findings such as XEN gel stent, Ex-PRESS device, long tube, trabeculectomy, anglesupported IOL. No adverse events related to gonioscopy or GS-1 examination were observed. Conclusions: AGP using the GS-1 is a reliable method for assessing apparent iris insertion and proved useful in detecting pathological and post-surgical ACA findings.
2018
Obiettivi: valutazione della goniofotografia automatica (AGP) - immagini a colori dell’angolo camerulare (ACA) - nell’identificazione delle strutture anatomiche angolari e dei reperti patologici o post-chirurgici. Metodi: lo strumento GS-1 (Gonioscope GS-1; NIDEK Technologies Srl) è stato utilizzato per acquisire immagini a 360 gradi dell’ACA in entrambi gli occhi di 50 pazienti. L’inserzione apparente dell’iride per ogni quadrante ed eventuali altri reperti dell’ACA sono stati registrati. Tutti i pazienti sono stati sottoposti anche alla gonioscopia standard e l’angolo è stato valutato utilizzando la classificazione di Spaeth. La concordanza tra i due strumenti è stata determinata utilizzando il coefficiente di Cohen's pesato linearmente κ (KW). Risultati: dei 400 quadranti fotografati, 339 (84.8%) erano di qualità sufficiente per le analisi. L’età media (DS) era 56 (14.2) anni e 29 (58%) erano donne. La concordanza tra AGP con GS-1 e gonioscopia standard è risultata essere buona (KW, 0.65; 95% CI, 0.51 to 0.78). Non ci sono state differenze tra AGP e gonioscopia circa l’identificazione di reperti anormali quali: XEN gel stent, dispositivo Ex-PRESS, tubicino di impianto drenante, trabeculectomia, IOL a supporto angolare. Nessun evento avverso relativo alla gonioscopia o al GS-1 è stato osservato.
File in questo prodotto:
File Dimensione Formato  
18_01_02-Clinical-evaluation-of-a-novel-device-for-anterior-chamber-angle-photography.pdf

accesso chiuso

Descrizione: file lavoro
Tipologia: Documento in versione editoriale
Dimensione 1.08 MB
Formato Adobe PDF
1.08 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11567/915985
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact