This in vitro study analyzed the reverse torque (RevT) of abutment screws following different cleaning treatments. A convenience sample of 50 customized titanium abutment screw complexes was divided into five groups: cleaning by steam (control group), cleaning by Argon plasma (test groups 1 and 2 [with chlorhexidine gel]), and replacement of old screws with new ones (test groups 3 and 4 [with chlorhexidine gel]). Abutments were screwed onto implants and tested for RevT. The RevT of the test groups was significantly higher than that of the control group. No statistically significant difference between test groups was noted except between groups 2 and 3.

Implant abutment screw reverse torque values before and after plasma cleaning

Canullo L;
2013-01-01

Abstract

This in vitro study analyzed the reverse torque (RevT) of abutment screws following different cleaning treatments. A convenience sample of 50 customized titanium abutment screw complexes was divided into five groups: cleaning by steam (control group), cleaning by Argon plasma (test groups 1 and 2 [with chlorhexidine gel]), and replacement of old screws with new ones (test groups 3 and 4 [with chlorhexidine gel]). Abutments were screwed onto implants and tested for RevT. The RevT of the test groups was significantly higher than that of the control group. No statistically significant difference between test groups was noted except between groups 2 and 3.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11567/1102295
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 19
  • ???jsp.display-item.citation.isi??? 17
social impact